Design and specifications

  • Collection Optics: Primary Detection Dichroic: 60 x 40 x 1
  • : Collection lens diameter: 45 mm
  • : Maximum collection angle: ~±8° assuming 20mm objective back aperture
  • : IR Blocking Filter: 680nm Shortpass 50mm Ø
  • Emission Filters: Emission filter: 32mm Ø - one for each detector
  • : Short wavelength perpendicular to the incident light, longer wavelength on the incident direction
  • : Blocking filter: 32mm Ø for stimulation wavelength
  • Photomultiplier Tubes: Two locally mounted detectors of the same type: GaAsP protected photomultiplier tube (H10770PA-40) or GaAsP photomultiplier tube with gating function (H11706P-40)
  • : Signal output Connector: SMA
  • : Gate input connector (gated GaAsP only): BNC
  • Protection Circuitry: Protected GaASP: Integrated protection circuitry to trip PMT when overload detected
  • : Gated GaAsP: Voltage gating functionality to avoid overload situations
  • : All systems: Ambient light detector
  • Objective and Nosepiece Compatibility: Objectives: M32, M27, M25, RMS threaded objectives (<20mm back aperture)
  • : Nosepieces: Ships with single objective changer with 2 locking sliders
  • : Compatible with: WI-SRE3 adapter (RMS only) or standard/DIC objective sliders
  • Preamplifier Bandwidth: Options include: 20 MHz (resonant system), 1 MHz (galvo), or variable bandwidth all with low noise linear power supply
  • Output Connection: BNC
  • PMT Voltage Control: Manual or Software Control
  • PMT Protection: Integrated protection circuitry in H10771P-40 only. For all PMT options, a remote ambient light sensor is used