Hamamatsu Orca Flash 4.0 V3 Technical Data
Hamamatsu Orca Flash 4.0 V3

Hamamatsu Orca Flash 4.0 V3

Specification

  • Quantum efficiency: 82 % (peak QE: @ 560 nm)
  • Effective no. of pixels: 2048 (H)×2048 (V)
  • Cell size: 6.5 μm×6.5 μm
  • Effective area: 13.312 mm×13.312 mm
  • Full well capacity: 30 000 electrons (typ.)
  • Readout speed: 100 frames/s (Full resolution, standard scan, Camera Link)
  • : 40 frames/s (Full resolution, standard scan, USB 3.0, 16 bit)
  • : 53 frames/s (Full resolution, standard scan, USB 3.0, 12 bit)
  • : 80 frames/s (Full resolution, standard scan, USB 3.0, 8 bit)
  • Readout noise: Standard scan (at 100 frames/s, typ.):1.6 electrons rms (1.0 electrons median)
  • : Slow scan (at 30 frames/s, typ.): 1.4 electrons rms (0.8 electrons median)
  • Cooling method: Peltier cooling
  • Cooling temperature: Forced air (Ambient at +20 ℃): -10 ℃
  • : Water (+20 ℃): -10 ℃
  • : Water (+15 ℃): -30 ℃
  • Dark current: 0.06 electrons/pixel/s (Air Cooled to -10° C) (typ.)
  • : 0.06 electrons/pixel/s (Water Cooled to -10° C) (typ.)
  • : 0.006 electrons/pixel/s (Water Cooled to -30° C) (typ.)
  • Dynamic range: 37 000:1 (typ.)
  • Interface: Camera Link / USB 3.0
  • A/D converter: 16 bit / 12 bit / 8 bit
  • Readout modes: Normal Area, Lightsheet, W-VIEW Mode, Dual Lightsheet
  • Binning: 2x2, 4x4
  • Master Pulse Generator (Pulse Modes): Internal Sync, Start Trigger, Burst
  • Master Pulse Generator (Pulse Interval in 1 μs increments): 10 μs to 10 s
  • Dark Signal Non-Uniformity (DSNU): 0.3 electrons rms
  • Photo Response Non-Uniformity (PRNU) at half level of full light range (15,000 electrons): 0.06 % rms
  • Photo Response Non-Uniformity (PRNU) at low light level (700 electrons): 0.3 % rms
  • Linearity error, full light range (EMVA 1288 standard): 0.5 %
  • Linearity error, low light range (< 500 electrons signal): 0.2 % / Less than approx. 1 electron absolute error

Contact Form

Contact us

*为必须填写内容

选择感兴趣领域